Service

Input/Output Description Analog Inputs (3) Selectable full-scale: 0 to 1V, 0 to 5V, 0 to 10V; 100 kohms input impedance; Bandwidth = 5Hz; Resolution: 12 bits ±1 LSB of full scale Analog Output (1) Selectable full-scale range:0 to 1V, 0 to 10V; Bandwidth = 10 Hz; Resolution = 8 bits; ECG Monitor Input Input voltage range 0 to 10V full scale; Channel Bandwith = 0.5 to 40 Hz; Resolution = ±1 BPM; Accuracy = ± 1 BPM; Range = 30-250 BPM; Digital I/O (2) RS232 programmable serial ports; Maximum data rate = 19.2 kilbaud USB Port One V1.1-compatible type A connector.

Edwards Lifesciences Vigilance Ii Service Manual

Input/Output Description Analog Inputs (3) Selectable full-scale: 0 to 1V, 0 to 5V, 0 to 10V; 100 kohms input impedance; Bandwidth = 5Hz; Resolution: 12 bits ±1 LSB of full scale Analog Output (1) Selectable full-scale range:0 to 1V, 0 to 10V; Bandwidth = 10 Hz; Resolution = 8 bits; ECG Monitor Input Input voltage range 0 to 10V full scale; Channel Bandwith = 0.5 to 40 Hz; Resolution = ±1 BPM; Accuracy = ± 1 BPM; Range = 30-250 BPM; Digital I/O (2) RS232 programmable serial ports; Maximum data rate = 19.2 kilbaud USB Port One V1.1-compatible type A connector.